The Burdick Atria 3100 ECG

The Atria 3100 is compact yet powerful. Our technology assures high quality and accuracy while giving protocol choices, helping eliminate transcription errors, and improving workflow.

  • Reduce typing mistakes and save time by downloading patient information from your electronic medical record (EMR) system.
  • Quickly review test results – reports can be viewed, edited, and confirmed right from the patient’s chart.
  • Choose to print beat-to-beat R-R intervals on reports.
  • Eliminate the hassle of scanning and filing hard copies by transmitting test results directly to your EMR system.
  • In short, the Atria 3100 makes your work life easier while delivering reliable performance. This might be a good time to transition to a device that does so much for you.

    TECHNICAL SPECIFICATIONS

    DATA STORAGE

    50 records

    NETWORK CONNECTION

    Wired 10/100 MBPS Ethernet via RJ45
    Wireless 802.11g (compatible with 802.11b networks), security options includingWEP (64/128 bit), WPA / WPA2 (Personal and Enterprise, including
    AES), EAP-TLS, PEAP/MSCHAPv2, LEAP, RADIUS servers (including
    Microsoft IAS, Cisco ACS and Cisco Aironet 1200), roaming across
    WAPs of various configurations and support for mixed 802.11 / Ethernet
    networks.

    POWER REQUIREMENTS

    AC operation 115/230 V AC +/-10%, 50/60 Hz at external power source
    Battery operation 14.4 V NiMH rechargeable battery pack
    Battery duration 300 pages continuous printing without communications enabled
    PRINTOUT Printout device 216 mm thermal dot array
    Paper dimensions 8.5 in x 11 in (US letter) or 210 mm x 300 mm (A4)
    Paper type Thermal sensitive (Burdick AssuranceR or Heartline™ paper recommended)
    Chart speeds 12.5, 25, 50 mm/sec
    Gain 5, 10, 20 mm/mV chest or limb (may be split)
    Printout formats 3-, 4-, 6-, or 12-channel; additional rhythm formats

    ACQUISITION

    Meets IEC 60601-2-51
    Lead selection I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V4R; supports Cabrera
    and alternate chest lead (chest lead selection V2R thru V9R + V7,V8,V9)
    Measurement (standard), Interpretation (optional)
    Diagnosis, measurements and reason statements based on five demo
    graphic criteria
    Pacemaker display capability Meets or exceeds ANSI/AAMI EC11-1991
    Modes Automatic, rhythm or manual
    Frequency response Meets or exceeds ANSI/AAMI EC11-1991 standard
    Input impedance Meets or exceeds ANSI/AAMI EC11-1991 standard
    Electrode offset tolerance +/- 300 mV
    A/D conversion 8,000 samples/second, 2.5 μV LSB RTI
    Artifact filter response 40 Hz, -3db
    Data storage resolution 500 samples/second, 2.5 μV LSB RTI

    SAFETY

    Leakage current: Patient <10 μA, chassis <100 μA;
    Defibrillation protection: to 5000 V, 360J safety STANDARDS ANSI/AAMI EC11-1991 UL 60601-1
    CAN/CSA C22.2 no. 0-M91 EN 60601
    CAN/CSA C22.2 no. 601.1S1 EN 60601-1-2:2001
    CAN/CSA C22.2 no. 601.2.25, Amend. 1 EN 60601-2-51:2003
    CAN/USA C22.2 no. 601.1-M90 EN 60601-1-4:2000
    DISPLAY Backlit 2 x 40 character LCD

    KEYBOARD

    Full alphanumeric keypad plus designated quick keys

    ENVIRONMENTAL

    Operating temperature 50°F to 104°F (10°C to 40°C)
    Operating relative humidity 20% to 90%, non-condensing
    Operating atmospheric pressure 1060 hPa to 700 hPa (-500 ft to 10,000 ft reference to sea level)
    Storage temperature -4°F to 115°F (-20°C to 45°C)
    Storage relative humidity 10% to 90%, non-condensing
    Storage atmospheric pressure 1060 hPa to 190 hPa (-500 ft to 40,000 ft reference to sea level)

    DIMENSIONS (H x D x W)

    5.5 in x 13.125 in x 15.0 in (14.0 cm x 33.4 cm x 38.1 cm)
    weight 11 lb (5 kg) (including external power supply)

    WARRANTY

    3 years with return of warranty card

    For more information on this product and the options available please contact our sales team at sales@mesm.co.uk